![]() Specification CAUTION: Low pO2 levels can influence the linearity of glucose measurements, and can therefore result in falsely low glucose results. This information is already provided in the ABL80 FLEX Reference Manual, chapter 6. Instructions to user Please place this document in the binder of your manual.īrief Overview of the changes Changes Description To increase operator awareness, the specifications listed here regarding glucose linearity in the ABL80 FLEX analyzer will be added to the operators manual. The following table outlines the glucose linearity as a function of the pO2. Low pO2 levels can influence the linearity of the glucose sensor. Operator Abl 80 Flex Radiometer Abl80 Flex Software Included In.